Infusion syringe pump
Sino SN-50C6
Single channel syringe pump is a the most common see and popular iv syringe driver pump among all models. It features stable quality, easy to operate, and automatic calibration and most importantly accurate injection. Various modes support include Time mode, Speed mode, body weight mode and drug library mode. The model with R is machine with smart WIFI function.
The SN infusion syringe pump is designed for accurate dosing of drugs administered to a patient. This type of pump can be used to equip an anesthesiologist's workplace in the operating room, intensive care and resuscitation units, premature infants, neonatal intensive care units, chemotherapy and other places where precise dosage of drugs is used per unit of time. This device allows titration of drugs for anesthesia, vasoactive and inotropic drugs, parenteral nutrition, chemotherapy and other medicines.
Technical characteristics
Syringes | 5, 10, 20, 30, 50 ml of calibrated syringes | |
Flow rate | 0.1–1500 ml/h (specified value 0.1 999.9 ml/h, 0.1 ml/h step) | 50 ml syringe |
0.1–900 ml/h (0.1 ml/h step) | 30 ml syringe | |
0.1–600 ml/h (0.1 ml/h step) | 20 ml syringe | |
0.1–400 ml/h (0.1 ml/h step) | 10 ml syringe | |
Purge rate | 1500 ml/h | 50 ml syringe |
900 ml/h | 30 ml syringe | |
600 ml/h | 20 ml syringe | |
400 ml/h | 10 ml syringe | |
Аlarm | Occlusion, Nearly empty, Finish, Syringe dislocated, Plunger/Clutch disengaged, Battery low, Battery exhausted, Power cable disconnected, Flow rate over limit, Volume over limit, System error, Non-operation time over | |
Display range of volume delivered | 0.1 ~ 9999 ml (0.1 ~ 999.9 ml,0.1 ml step) (1000 ~ 9999 ml, 1 ml step) | |
Setting of delivery limit | 0.1 ~ 9999 ml (0.1 ~ 999.9 ml,0.1 ml step) (1000 ~ 9999 ml, 1 ml step) | |
Occlusion detection pressure | H: 800±200 mmHg (106.7±26.7 kPa) or 1.09±0.27±kgf/cm2 C: 500±100 mmHg (66.7±13.3 kPa) or 0.68±0.13±kgf/ cm2 L: 300±100 mmHg (40.7±13.3 kPa) or 0.41±0.13±kgf/ cm2 | |
Accuracy of flow rate (Nominal value) | Accuracy including syringe: with in ±2% Mechanical accuracy: with in ±1% | |
Power | AC110-220V± 10%,50/60 Hz±1 Hz Inner NiMH battery: DC12 V Continuous operation time of battery: over 6 hours for one channel, over 4 hours for dual channel when battery is recharged for 12 hours, at flow rate of 5 ml/h. | |
Operation conditions | Ambient temperature: 5 to 40 °C Relative humidity: 20% to 80%/td> | |
Storage conditions | Ambient temperature: 5 to 40 °C Relative humidity: ≤95% | |
Electric shock protection/class | Class I and internal power supply: Type CF, Continuous operation | |
Waterproof level | IPX 4 | |
Weight /class | 2.25 kg (with clamp base)/td> | |
Dimensions | 306 mm (Width) x 135 mm (Height) x 127 mm (Depth) |
Function
Calibrating syringe automatically | Fast rate control |
Identifying syringe size automatically | «Purge» button activate safe protection |
3-level occlusion alarm limits setting | Flow rate setting button lock |
Delivered volume inquiry | Abundant optic and sound alarm |
Limits setting | AC and internal power indication |
Automatically reducing impact dose release after occlusion | Standard RS232 interface |
KVO flow rate (0.1–1.0 ml/h Adjustable) | 1000 history records |
Wireless communication (SN-50C66R) | Multi-direction clamp for either horizontal or vertical fixation |