Automatic humidifier MG 3000 is designed to humidify and heat the respiratory mixture delivered to a patient during positive airway pressure respiratory support though an endotracheal tube, nasal cannulas or a face mask with availability of constant flow.

Heating and humidifying of gas is provided by the flow passing above the surface of the heated water, the constant temperature in the breathing circuit tubes is maintained by one or two heating elements located in it.

Automatic humidifier MG 3000 has the following features:

  • temperature in the circuit is controlled by two probes located at the output of a humidifier chamber and at the breathing circuit proximal end level (at the endotracheal tube)
  • two-point temperature control function ensures that the respiratory mixture delivered to the patient has appropriate level of heating
  • temperature level of the respiratory mixture by using Y-type connector can be set automatically
  • by using the button on the front panel, it can be possible to select the option of respiratory support (invasive or non-invasive)
  • temperature of the respiratory mixture is displayed on the LED display

Main characteristics

Using breathing circuit of different manufactories
Using non-heated breathing circuit
Using heated breathing circuit
Displaying all possible temperature parameters in the view mode
Setting a constant displaying one of the temperature parameters
Automatic and manual keyboard lock
Alarm temporary switch off (2 minutes)
Emergency protection of the heating plate from overheating
Invasive / non-invasive ventilation mode
Temperature control in «heated breathing circuit mode» (invasive mode)
airway≥38,0°C, ≤40°C
humidifier chamber≥37°C
Temperature control in «heated breathing circuit mode» (non-invasive mode)
airway≥34°C
humidifier chamber≥31°C
Temperature control in «non-heated breathing circuit mode» (invasive mode)
airway≥37°C
humidifier chamber≤66°С
Temperature control in «non-heated breathing circuit mode» (non-invasive mode)
airway≥31°C
humidifier chamber≤66°С
Automatic adjustment of the heating level of the inspiratory line
4-bit 60 segment display
Temperature displayed range (no less)5,0–80,0°C
Accuracy of temperature displayed range (temperature range of 25,0 to 45,0°C, no more)0,3°С
Water output at 37,0°C temperature (no less)44 mgH2O/l
Connector for heater circuit
Connector for temperature probe
Displayed parameters
Set temperature in airway (°C)
Set temperature in humidifier chamber (°C)
Actual temperature in airway
Actual temperature in humidifier chamber
Actual temperature of heating plate
System error code
Alarm systems
Displaying alarm signals on the front panel
Temperature probe is not connected, temperature probe is faulty
Inhalation line heater breathing circuit is not connected to the main unit, inhalation line heater breathing circuit is faulty
Exhalation line heater breathing circuit is not connected to the main unit, exhalation line heater breathing circuit is faulty
Temperature in humidifier chamber >41,0°C
Temperature in humidifier chamber <26,0°C for 10 minutes
Temperature in airway >41,0°C
Heated circuit: temperature <29.5°C for 10 minutes and <34,5°C for 60 minutes
Non-heated circuit: temperature <29.5°C for 10 minutes (invasive mode) and <26,0°C for 10 minutes after heating (non-invasive mode)
See user manual (microprocessor malfunction, system malfunction)
Alarm signal (audible)
Light indication alarm
Failure encoding system