Main page
Request for a commercial offer
Send message Order a callback Application for equipment
 

Infusion syringe pump
SN-S

  • Infusion syringe pump SN-S

Send a request
to clarify the cost,
terms and delivery terms Infusion syringe pump SN-S


Update
 

 

The SN infusion pump is designed for accurate dosing of drugs administered to a patient. This type of pump can be used to equip an anesthesiologist's workplace in the operating room, intensive care and resuscitation units, chemotherapy, cardiology and other places where precise dosage of drugs is used per unit of time. This device allows titration of drugs for anesthesia, vasoactive and inotropic drugs, parenteral nutrition, chemotherapy and other medicines. The infusion pump is characterized by high accuracy of drug infusion, stability of the rate of administration, and the possibility of micro dosing of administered drugs.

Technical characteristics

Infusion Mode SN-S1A: Fully extruded type, Speed Mode, Time Mode, Body-weight Mode, Drip Mode
SN-S1: Fully extruded type, Speed Mode, Time Mode, Body-weight Mode, Micro Mode, Drip Mode, Sequence Mode, Gradient Mode
SN-S2A: Half extruded type, Speed Mode, Time Mode, Body-weight Mode, Drip Mode
SN-S2: Half extruded type, Speed Mode, Time Mode, Body-weight Mode, Drip Mode, Sequence Mode, Gradient Mode, Micro Mode
Infusion Set 20d/ml, 60d/ml
Infusion type SN-S1 Series is Fully extruded peristaltic infusion.
SN-S2 Series is half extruded peristaltic infusion.
Accurancy of flow rate ±5% or 3d/min
Flow rate 20d/ml IV set flow rate range
Micro Mode: 0.1ml/h–100.0 ml/h
Speed Mode: 0.1ml/h–1500.0 ml/h
Drip Mode: 1 d/min–400.0 d/min

60d/ml IV set flow rate range
Micro Mode: 0.1ml/h–150.0 ml/h
Speed Mode: 0.1ml/h–200.0 ml/h
Drip Mode: 1 d/min–200.0 d/min
Present Volume 0.1–9999.99 ml, increment 0.01 ml
Delivered Volume 0.1–9999.99 ml, increment 0.01 ml
Аlarms Operation Timeout Alarm, Nearly Empty Alarm, Non-calibration of infuion set, Occlusion, Bubble, Door open, Finish, KVO Finish, Drip sensor abnormal, No AC Power, Battery Low, Battery Exhausted, System Erroor, System Power Error, Over Limit, Drip clamp not closed, Not installed battery, Infusion speed abnormal.
KVO rate 0.1–5 ml/h
BOLUS 5–1500 ml/h
Bubble Detector Bubble Detector in 8 level Default: Level 3:
Level 1: 25 ul
Level 2: 50 ul
Level 3: 100 ul
Level 4: 200 ul
Level 5: 300 ul
Level 6: 400 ul
Level 7: 500 ul
Level 8: 800 ul
Occlusion Detector Pressure The alarm threshold is between 13.3 kPa-120kPa with 12 levels devided linearly;
The default level is 5 in Micro mode;
The default level is 8 in other infusion mode
History More than 2000 pcs
Special Function Supporting pump stackable with working station, Relay infusion, Keypad Lock, Adjusting rate on infusion, Release pressure automatically;
Peristaltic pump can be taken off to clean;
Anti-retract detector, Drug library
Infusion Mode SN-S1A: Fully extruded type, Speed Mode, Drip Mode, Body-weight Mode, Time Mode
SN-S1: Fully extruded type, Speed Mode, Drip Mode, Body-weight Mode, Time Mode, Sequence Mode, Micro Mode, Gradient Mode
SN-S2A: Half extruded type, Speed Mode, Drip Mode, Body-weight Mode, Time Mode
SN-S2: Half extruded type, Speed Mode, Drip Mode, Body-weight Mode, Time Mode, Sequence Mode, Micro Mode, Gradient Mode
Communication RS485, WIFI
Protection Class Class II, Type CF, IP33
Dimension/ Weight 287x168x82 (mm), 2.1 KG
Interface DC Power interface, General Communication Interface (Nurse call, RS485)
Power a. AC Adapter
Input: 100–240 Va. C, 50/60 Hz, 40VA
Output: 15Vd.c,2.0A

b. DC Power: DC 12V

c. Battery: 7.4 V Li, 2600 mAh or 5200mAh
Rechaining time: Approx 8 hours when power on; Approx 4 hours when power off
Battery operating time:
10 hours @5200mAh at 5ml/h
5 hours @2600mAh at 5ml/h
Environment Operation condition:
Ambient Temperature: +5°C–40°C
Relative Humidity: 15%–95% (Non-condensed)
Pressure: 57.0 kPa–106 kPa, ≤ 3000 m Altitude

Storage condition:
Ambient Temperature: -40°C...+70°C,
Relative Humidity: 10%–98% (Non-condensed)
Pressure: 50.0 kPa–106 kPa